START OF THE ELPROVIEWER
2
2.1.5.1 MKT activation energy
The set standard value can only be changed if "Global activation energy" is selected.
In mode: Assessment - join the calculations are only carried out with one global value for
the activation energy. If different values are used in the PDF reports, a global value must
be specified.
2.1.5.2 Statistics
Logging with missing measurement values (gap) Reasons for missing measurement
values
- LIBERO with temporarily missing or short-circuited external sensor.
- Applications outside the device specifications.
- Only 1 measurement value n.c. or s.e.
Gaps of only 1 measurement value are automatically closed.
Statistics calculation if
measurement values
are missing
If there are gaps in the PDF report between the individual measurement values, there are
the following options:
- Ignore missing values
The statistics and the MKT are only calculated with the available measurement values.
This behavior corresponds to the LIBERO Tx. For the LIBERO Cx this setting can be
used to calculate the statistics and the MKT for existing data.
- Do not calculate the statistics
This behavior corresponds to the LIBERO Cx. The statistics and the MKT are not calcu-
lated.
Handling missing peri-
ods between two
curves (Join)
7.3 Join - Missing measurement values
For rounding reasons the accuracy of the MKT calculation is 0.5 K
n.c. Not connected
s.e. Sensor error