18 - EN
elproVIEWER
SI3001Ea
BEFORE YOU START
1
1.2.2 Type: LIBERO Cx
Housing elements 1. Protection cap
2. USB connector
3. Display; a not configured LIBERO Cx shows "ConF".
4. Start/Stop key
5. Exp:XX/YYYY
Expiry date: The LIBERO Cx should not be used after this date, irrespective of the
battery status.
Device ID: 7100 0000 0000
Display 1. Run
Flashes if it the LIBERO Cx is active; delay or data recording
2. Battery down
Indicates a low battery voltage. Data access via the USB port is still possible.
3. Profile
The displayed value corresponds to the profile ID
4. Alarm indicator
The alarm indicator may be triggered independently of defined alarm criteria by device
error or by using the device out of measurement range
5. Temperature Unit
6. Measurement value or information
1. Alarm status: Alarm
8.1.3 Extended alarm functions LIBERO Cx
Configuration -
Start - Logging -
Assessment
8 LIBERO PDF report
MKT calculation If there are n.c. values (no sensor connected) or s.e. values (sensor short-circuit) in the
measured value table, the LIBERO Cx does not calculate any statistics or any MKT.
This can be changed in the calculation options of the elproVIEWER.
2.1.5 Calculation.
Comentários a estes Manuais